Automated instrument designed to determine visual photosensitivity thresholds

  • Aguilar M
  • Gonzalez A
  • Rowaan C
  • et al.
18Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.
Get full text

Abstract

© 2018 Optical Society of America. The Ocular Photosensitivity Analyzer (OPA), a new automated instrument to quantify the visual photosensitivity thresholds (VPT) in healthy and light sensitive subjects, is described. The OPA generates light stimuli of varying intensities utilizing unequal ascending and descending steps to yield the VPT. The performance of the OPA was evaluated in healthy subjects, as well as light sensitive subjects with achromatopsia or traumatic brain injury (TBI). VPT in healthy, achromatopsia, and TBI subjects were 3.2 ± 0.6 log lux, 0.5 ± 0.5 log lux, and 0.4 ± 0.6 log lux, respectively. Light sensitive subjects manifested significantly lower VPT compared to healthy subjects. Longitudinal analysis revealed that the OPA reliably measured VPT in healthy subjects.

Cite

CITATION STYLE

APA

Aguilar, M. C., Gonzalez, A., Rowaan, C., de Freitas, C., Alawa, K. A., Durkee, H., … Parel, J.-M. A. (2018). Automated instrument designed to determine visual photosensitivity thresholds. Biomedical Optics Express, 9(11), 5583. https://doi.org/10.1364/boe.9.005583