Abstract
Purpose: To optimize the accuracy and precision of T2 measurements using the standard Carr-Purcell-Meiboom-Gill (CPMG) sequence. T2 values obtained with this technique are normally sensitive to imperfect refocusing due to the formation of unwanted stimulated echoes. Materials and Methods: Modifications are made to the refocusing slice selection width and the interleaving scheme. A widened refocusing slice improves the uniformity of the refocusing flip angle across the slice. A slow spin echo acquisition provided "gold standard" T2 values. Repeated T2 measurements in phantom and human studies are used to compare the accuracy and precision of the optimized and non-optimized CPMG implementations. Results: The required slice thickness ratio between refocusing and excitation slice widths is found to be 3:1 for typical optimized radiofrequency pulses. T2 values obtained using this optimized implementation more closely correspond to "gold standard" values. Repeated T2 measurements indicate significantly improved correspondence between data and model. A reduction in the fitting error of approximately 70% is demonstrated for phantoms. Conclusion: We demonstrate that a relatively simple change to the CPMG relaxometry sequence parameters from the default setup yields significant improvements in the accuracy and precision of T2 measurements. © 2006 Wiley-Liss, Inc.
Author supplied keywords
Cite
CITATION STYLE
Pell, G. S., Briellmann, R. S., Waites, A. B., Abbott, D. F., Lewis, D. P., & Jackson, G. D. (2006). Optimized clinical T2 relaxometry with a standard CPMG sequence. Journal of Magnetic Resonance Imaging, 23(2), 248–252. https://doi.org/10.1002/jmri.20490
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.