3D cones acquisition of human extremity imaging using a 1.5T superconducting magnet and an unshielded gradient coil set

4Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

Abstract

We developed ultrashort echo-time (UTE) imaging sequences with 3D Cones trajectories for a home-built compact MRI system using a 1.5T superconducting magnet and an unshielded gradient coil set. We achieved less than 7 min imaging time and obtained clear in vivo images of a human forearm with a TE of 0.4 ms. We concluded that UTE imaging using 3D Cones acquisition was successfully implemented in our 1.5T MRI system.

Cite

CITATION STYLE

APA

Setoi, A., & Kose, K. (2019). 3D cones acquisition of human extremity imaging using a 1.5T superconducting magnet and an unshielded gradient coil set. Magnetic Resonance in Medical Sciences. Japanese Society for Magnetic Resonance in Medicine. https://doi.org/10.2463/mrms.tn.2017-0170

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free