Easy retrieval of polyaxial tulip-head pedicle screws by "u" rod technique

4Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.

Abstract

The number of fusion surgeries increase each year which also increase the need for implant removal. In some cases, it can be extremely hard to remove a pedicle screw especially when there is a mismatch of the screw and the screwdriver. Also the screwdrivers can be contaminated during the operation, and this will cause a delay till the instruments are re-sterilized. There is a need for the removal of screws without special instruments. We describe a method for removing tulip-head polyaxial pedicle screws without special instruments. The screws are removed using an Allen key, a rod bender and a "U" shaped rod. We successfully removed 76 screws in 11 recent cases without any complications. The "U" rod technique is a simple and useful technique for the removal of tulip-head polyaxial screws. © 2009 Springer-Verlag.

Cite

CITATION STYLE

APA

Kose, K. C., Isik, C., Altinel, L., Ates, A., & Ozdemir, M. (2010). Easy retrieval of polyaxial tulip-head pedicle screws by “u” rod technique. European Spine Journal, 19(1), 144–146. https://doi.org/10.1007/s00586-009-1096-4