Purpose: Three-dimensional, time-resolved blood flow measurement (4D-flow) is a powerful research and clinical tool, but improved resolution and scan times are needed. Therefore, this study aims to (1) present a postprocessing framework for optimization-driven simulation-based flow imaging, called 4D-flow High-resolution Imaging with a priori Knowledge Incorporating the Navier-Stokes equations and the discontinuous Galerkin method (4D-flow HIKING), (2) investigate the framework in synthetic tests, (3) perform phantom validation using laser particle imaging velocimetry, and (4) demonstrate the use of the framework in vivo. Methods: An optimizing computational fluid dynamics solver including adjoint-based optimization was developed to fit computational fluid dynamics solutions to 4D-flow data. Synthetic tests were performed in 2D, and phantom validation was performed with pulsatile flow. Reference velocity data were acquired using particle imaging velocimetry, and 4D-flow data were acquired at 1.5 T. In vivo testing was performed on intracranial arteries in a healthy volunteer at 7 T, with 2D flow as the reference. Results: Synthetic tests showed low error (0.4%-0.7%). Phantom validation showed improved agreement with laser particle imaging velocimetry compared with input 4D-flow in the horizontal (mean −0.05 vs −1.11 cm/s, P
Author supplied keywords
Cite
CITATION STYLE
Töger, J., Zahr, M. J., Aristokleous, N., Markenroth Bloch, K., Carlsson, M., & Persson, P. O. (2020). Blood flow imaging by optimal matching of computational fluid dynamics to 4D-flow data. Magnetic Resonance in Medicine, 84(4), 2231–2245. https://doi.org/10.1002/mrm.28269
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.