Characterization of cardiac acoustic biomarkers in patients with heart failure

9Citations
Citations of this article
22Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: The wearable cardioverter defibrillator (WCD) records electrocardiograms and cardiohemic vibrations that can be algorithmically combined to provide cardiac acoustic biomarkers (CABs). We characterized CAB variability, diurnal variations, and changes over time among heart failure patients. Methods: Wearable cardioverter defibrillator heart failure patients who had CAB recordings from March 2015 to July 2017 were included. CAB parameters included: electromechanical activation time (EMAT), EMATc (EMAT/RR interval), left ventricular systolic time (LVST), LVSTc (LVST/RR interval), S3 and S4 strengths, and systolic dysfunction index (SDI). Descriptive statistics, correlation analysis, and analysis of variance were used to report temporal and clinical associations. Results: One thousand and sixty-six WCD patients met the study criteria. Diastolic CAB parameters showed significantly greater intra-subject variability than systolic CAB parameters (>29% vs. <15%, p