Note: Rapid offset reduction of impedance bridges taking into account instrumental damping and phase shifting

1Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The sensitivity of an imperfectly balanced impedance bridge is limited by the remaining offset voltage. Here, we present a procedure for offset reduction in impedance measurements using a lock-in amplifier, by applying a complex compensating voltage external to the bridge. This procedure takes into account instrumental damping and phase shifting, which generally occur at the high end of the operational frequency range. Measurements demonstrate that the output of the circuit rapidly converges to the instrumentally limited noise at any frequency. © 2013 American Institute of Physics.