Abstract
Objective: We verified a magnetic bead-based, simple, and fast method for circulating cell-free DNA (cfDNA) extraction from whole blood samples(CEWB) and characterised its utility in non-invasive prenatal testing (NIPT). Method: We extracted cfDNA from both plasma and whole blood of the patients using CEWB and compared it to that extracted using a Qiagen extraction kit; droplet digital polymerase chain reaction test was used to calculate the fragment size bias. In all, 304 samples were used for NIPT. Results: The CEWB group (mean ± standard deviation [SD]: 4.34 ± 0.41 ng/ml plasma) reported less DNA weight yield than the Qiagen group (4.90 ± 0.50 ng/ml plasma). There was no significant difference between the CEWB group and the Qiagen group in the gene fragments (136 bp: p = 0.064 and 420 bp: p = 0.534). In a parallel cohort study to characterise the utility of the CEWB method in NIPT, the treatment group extracted by CEWB showed a sensitivity of 100%, a specificity of 99.65%, and a positive predictive value of 95%. Conclusions: This study demonstrated that CEWB achieves an acceptable yield of DNA without contamination from genomic DNA. Subsequent clinical experiments in a parallel cohort indicated its utility for NIPT.
Cite
CITATION STYLE
Zhong, H., Zeng, L., Tao, M., Ye, Y., Wang, Y., Hou, L., … Tang, M. (2022). A novel method for extracting circulating cell-free DNA from whole blood samples and its utility in the non-invasive prenatal test. Prenatal Diagnosis, 42(9), 1173–1181. https://doi.org/10.1002/pd.6212