A new calibration setup for lock-in amplifiers in the low frequency range and its validation in a bilateral comparison

10Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.
Get full text

Abstract

This paper addresses the calibration of lock-in amplifiers in the low frequency range. A simple but effective calibration setup implemented at INRIM is described. This includes a stable low-voltage source, composed of a digital sine wave generator and a cascade of two voltage dividers, one inductive and one resistive. The magnitude error is the lock-in magnitude error, which can be determined with an uncertainty of the order of 10-4 or better, that is, two orders of magnitude lower than the typical lock-in accuracy specification. The system can operate with an input voltage from 0.1 µV to 10 µV and a frequency from a few hertz to about 1 kHz. The performances of the calibration setup were evaluated at 1 µV and 10 μV at 20 Hz and 103 Hz and compared with those of another existing setup, implemented at METAS, through a direct comparison in the framework of an EURAMET research project.

Cite

CITATION STYLE

APA

Cultrera, A., Corminboeuf, D., D’Elia, V., Tran, N. T. M., Callegaro, L., & Ortolano, M. (2021). A new calibration setup for lock-in amplifiers in the low frequency range and its validation in a bilateral comparison. Metrologia, 58(2). https://doi.org/10.1088/1681-7575/abdae0