Simultaneous determination by Gas chromatography of lindane and carbaryl in combined formulations

4Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A gas chromatographic method developed and validated for simultaneous determination of lindane and carbaryl for quality evaluation of lindane-carbaryl granule, using a glass capillary HP5 column (30 m × 0.32 mm; 0.25 μm), temperature programming with flame ionization detector and dibutylphthalate as an internal standard. The calibration graphs were found linear in the concentration range of 1 μg/mL to 1000 μg/mL for both lindane and carbaryl with correlation coefficient of 0.999 and 0.999 respectively and co-efficient of variation for intra-day and inter-day repeatability studies at different concentration levels was found to be less than 2%. The accuracy of method ranges between 98.5% to 100.8%. Specificity and robustness were also within the acceptable range. The method is highly sensitive with LOD and LOQ as 0.5 and 2 μg/mL for lindane and carbaryl respectively. The method has been tried on several formulations of lindane-carbaryl granules for quality control and has been found to be applicable.

Cite

CITATION STYLE

APA

Sen, I., Shandil, A., Aggarwal, M., & Khandal, R. K. (2011). Simultaneous determination by Gas chromatography of lindane and carbaryl in combined formulations. E-Journal of Chemistry, 8(1), 391–399. https://doi.org/10.1155/2011/195495