Immunoradiometric assay of atrial natriuretic peptide in unextracted plasma

10Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We have developed and validated a two-site liquid-phase immunoradiometric assay (IRMA) of atrial natriuretic peptide (ANP) in unextracted human plasma. Both radiolabeled rabbit anti-ANP IgQ and polyclonal mouse anti-ANP must bind to ANP for detection, and the assay is specific for peptides with both an intact C-terminus and a disulfide bridge. The assay sensitivity (detection limit) is 0.96 pmol/L, and the working range is 2.3-300 pmol/L, with the hook effect occurring above 500 pmol/L. Results for diluted plasma from normal subjects and from patients with renal failure paralleled the standard curve; analytical recovery of ANP added to such samples averaged 94%. The between- and within-assay CVs at 8 pmol/L were 10% and 5%, respectively. The assay is sufficiently sensitive and precise to detect the postural change in ANP concentrations in normal subjects.

Cite

CITATION STYLE

APA

Tattersall, J. E., Dawnay, A., McLean, C., & Cattell, W. R. (1990). Immunoradiometric assay of atrial natriuretic peptide in unextracted plasma. Clinical Chemistry, 36(6), 855–859. https://doi.org/10.1093/clinchem/36.6.855