Abstract
Aim: Conventional eyedropper-delivered volumes (25-50 μl) exceed the eye's usual tear-film volume (7 μl) and precorneal reservoir capacity, risking overflow and ocular/systemic complications. Piezoelectric high-precision microdosing may circumvent these limitations. Results & methodology: In this masked, nonrandomized, cross-over study, subjects (n = 12) underwent pupil dilation with topical phenylephrine (PE) administered by 32-μl eyedropper (2.5% or 10% formulation) and 8-μl electronic microdosing (10% formulation). Microdosing with PE-10% achieved comparable peak dilation as 10% eyedropper-delivery and superior dilation to 2.5% eyedropper-delivery (p = 0.009) at 75 min. Microdosing significantly reduced 20-min plasma PE levels versus PE10% eyedropper; neither treatment altered heart rate/blood pressure. Eye irritation occurred significantly less frequently with microdosing than PE10% eyedrops. Conclusion: Piezo-ejection PE microdosing achieves comparable biological effect as eyedropper dosing; reduced systemic absorption may decrease risk of systemic side effects.
Author supplied keywords
Cite
CITATION STYLE
Ianchulev, T., Weinreb, R., Tsai, J. C., Lin, S., & Pasquale, L. R. (2018). High-precision piezo-ejection ocular microdosing: Phase II study on local and systemic effects of topical phenylephrine. Therapeutic Delivery, 9(1), 17–27. https://doi.org/10.4155/tde-2017-0095
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.