Dextrostix-eyetone in the insulin hypoglycaemia test

12Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

The Ames Dextrostix-Eyetone system was evaluated for monitoring the blood glucose concentration during insulin-induced hypoglycaemia. The results agreed well with laboratory values for plasma glucose, obtained by an orthotoluidine method, and the method was practicable as a bedside technique. In two cases quick results obtained with the Eyetone enabled the insulin tolerance test to be interrupted to prevent severe hypoglycaemia before the clinical indications were obvious. The extra time and effort required were minimal, and its value seems to far outweigh the disadvantage of the extra work entailed. Nevertheless, care in using the system was important, and the operator must familiarise himself with the system before the most reliable results can be obtained. © 1977, British Medical Journal Publishing Group. All rights reserved.

Cite

CITATION STYLE

APA

Newall, R. G. (1977). Dextrostix-eyetone in the insulin hypoglycaemia test. British Medical Journal, 2(6080), 152–154. https://doi.org/10.1136/bmj.2.6080.152