Abstract
We have developed a robust current pattern for detection of intraventricular hemorrhage (IVH). In this study, the current pattern was applied on two realistic shaped neonatal head models and one head-shaped phantom. We found that a sensitivity matrix calculated from a spherical model gave us satisfactory reconstructions in terms of both image quality and quantification. Incorporating correct geometry information into the forward model improved image quality. However, it did not improve quantification accuracy. The results indicate that using a spherical matrix may be a more practical choice for monitoring IVH volumes in neonates. © 2010 IOP Publishing Ltd.
Cite
CITATION STYLE
Tang, T., & Sadleir, R. (2010). Quantification of intraventricular hemorrhage is consistent using a spherical sensitivity matrix. In Journal of Physics: Conference Series (Vol. 224). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/224/1/012064