Aims Echocardiographic optimization of atrioventricular (AV) and interventricular (VV) intervals in cardiac resynchronization therapy (CRT) is costly, time-consuming, and requires skill and expertise so is usually undertaken only in 'non-responder' patients. An algorithm in St Jude Medical CRT devices (QuickOpt™) claims to optimize these settings automatically. The aim of this study was to compare the two optimization techniques.Methods and resultsOptimization of AV and VV intervals was performed a month after CRT device implantation in 26 patients with heart failure, first by echocardiography then by QuickOpt. The left ventricular outflow tract (LVOT) velocity-time integral (VTI) was measured after optimization by each method. Agreement between the optimization methods was assessed by the Bland-Altman analysis and correlation by Pearson's correlation coefficient. There was good correlation between the LVOT VTI following optimization by both methods (R2 = 0.77, P < 0.001). However, agreement between the two methods was poor, with 15 of 26 and 10 of 26 patients having a >20 ms difference in the optimal AV and VV interval values, respectively. Left ventricular outflow tract VTI was significantly better (22 of 26 patients; P < 0.001) in patients optimized by echocardiography than by QuickOpt.ConclusionThere is a poor agreement in optimal AV and VV intervals determined by echocardiography and QuickOpt, with echocardiographic optimization giving a superior haemodynamic outcome.
Author supplied keywords
Cite
CITATION STYLE
Kamdar, R., Frain, E., Warburton, F., Richmond, L., Mullan, V., Berriman, T., … Schilling, R. (2010). A prospective comparison of echocardiography and device algorithms for atrioventricular and interventricular interval optimization in cardiac resynchronization therapy. Europace, 12(1), 84–91. https://doi.org/10.1093/europace/eup337
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.