Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement

116Citations
Citations of this article
80Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

BackgroundArterial pressure (AP) monitoring should be accurate, easy to use, free of risks, and ideally continuous. The continuous non-invasive arterial pressure (CNAP) device is non-invasive and provides continuous pressure readings. This study was performed to compare the agreement of CNAP and invasive AP monitoring. MethodsNinety patients undergoing surgery under general anaesthesia were enrolled. Invasive pressure monitoring was established at the radial artery. CNAP monitoring using a finger sensor recording was begun before induction of anaesthesia. Statistical analysis was conducted with the BlandAltman method for comparisons of repeated measures. ResultsWe obtained 16 843 valid pressure readings from 85 patients. Mean (sd) bias during maintenance of anaesthesia was: systolic AP: 4.2 (16.5) mm Hg; mean AP (MAP):-4.3 (10.4) mm Hg; and diastolic AP:-5.8 (6) mm Hg. The results of a subgroup analysis of patients who had a mean intra-arterial pressure of <70 mm Hg were as follows: systolic pressure:-0.3 (9.7) mm Hg; mean pressure:-6.8 (7.6) mm Hg; and diastolic pressure:-7.9 (7.2) mm Hg. Bias and percentage error during the induction period were greater in both the main and subgroup analyses, probably due to recalibration being omitted after induction. ConclusionsThe CNAP monitor showed an acceptable agreement and was interchangeable with invasive pressure monitoring for MAP during normotensive conditions. During induction of anaesthesia and when the AP was low, the agreement was less good and interchangeability was not achieved. These results suggest that CNAP is not statistically equivalent to invasive monitoring during all periods of anaesthesia but may be a useful additional AP monitor. © The Author [2012]. Published by Oxford University Press on behalf of the British Journal of Anaesthesia. All rights reserved.

Cite

CITATION STYLE

APA

Ilies, C., Bauer, M., Berg, P., Rosenberg, J., Hedderich, J., Bein, B., … Hanss, R. (2012). Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. British Journal of Anaesthesia, 108(2), 202–210. https://doi.org/10.1093/bja/aer394

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free