Abstract
Rolling circle amplification (RCA) and loop mediated isothermal amplification (LAMP) were combined to establish the rolling circle and loop mediated isothermal amplification (RC-LAMP) method for miRNA detection. With the participation of Bst 2.0 DNA Polymerase, the method enabled RCA and LAMP amplification to occur simultaneously without thermal cycling. The limit of detection of RC-LAMP was 500 amol/L, which is comparable to previously reported amplification strategies. Moreover, its upper limit of quantitation was higher and showed a stronger resistance to matrix interference. Therefore, it is possible to detect low concentrations of miRNA in samples by increasing the total RNA added. Owing to its facile detection mode and simple operation, this method has great potential in clinical sample detection.
Author supplied keywords
Cite
CITATION STYLE
Cao, Z., Jiang, X., Xiao, G., Xu, M., Liu, H., & Cai, S. (2021). Rolling circle and loop mediated isothermal amplification strategy for ultrasensitive miRNA detection. Separations, 8(10). https://doi.org/10.3390/separations8100166