Abstract
A new facility for primary shock force calibrations up to a force amplitude of 250 kN is presented. Shock forces are generated by the impact of two airborne cylindrical mass bodies. Two laser-Doppler interferometers simultaneously measure the bodies' dynamic motions on their common axis of motion. Traceability of force is realised by the determination of mass and acceleration. This paper is an extended version of the original contribution to the IMEKO 2010 conference in Pattaya, Thailand. © 2013 Elsevier Ltd. All rights reserved.
Author supplied keywords
Cite
CITATION STYLE
Kobusch, M., Bruns, T., Klaus, L., & Müller, M. (2013). The 250 kN primary shock force calibration device at PTB. Measurement: Journal of the International Measurement Confederation, 46(5), 1757–1761. https://doi.org/10.1016/j.measurement.2012.11.045