This report describes an asymptomatic patient with a DDDR pacemaker system (Inos Cls, Biotronik) that showed intermittent mode reversion to VOO and sometimes to VVI without a detectable source of electromagnetic interference and oversensing. Patients with this Biotronik model should be carefully assessed along with a surface ECG before placing the programming head.
Author supplied keywords
Cite
CITATION STYLE
APA
Erdogan, O., Altun, A., & Ozbay, G. (2001). Intermittent mode reversion to VOO of a dual chamber pacemaker model. PACE - Pacing and Clinical Electrophysiology, 24(11), 1704–1705. https://doi.org/10.1046/j.1460-9592.2001.01704.x