Abstract
Lensless microfluidic imaging with super-resolution processing has become a promising solution to miniaturize the conventional flow cytometer for point-of-care applications. The previous multi-frame super-resolution processing system can improve resolution but has limited cell flow rate and hence low throughput when capturing multiple subpixel-shifted cell images. This paper introduces a single-frame super-resolution processing with on-line machine-learning for contact images of cells. A corresponding contact-imaging based microfluidic cytometer prototype is demonstrated for cell recognition and counting. Compared with commercial flow cytometer, less than 8% error is observed for absolute number of microbeads; and 0.10 coefficient of variation is observed for cell-ratio of mixed RBC and HepG2 cells in solution. © 2014 Huang et al.
Cite
CITATION STYLE
Huang, X., Guo, J., Wang, X., Yan, M., Kang, Y., & Yu, H. (2014). A contact-imaging based microfluidic cytometer with machine-learning for single-frame super-resolution processing. PLoS ONE, 9(8). https://doi.org/10.1371/journal.pone.0104539