Abstract
Objective This study aimed to optimize computed tomography (CT) parameters for detecting ground glass opacity nodules (GGNs) using a computer-assisted detection (CAD) system and a lung cancer screening phantom. Methods A lung cancer screening phantom containing 15 artificial GGNs (-630 Hounsfield unit [HU], 2–10 mm) in the left lung was examined with a CT scanner. Three tube voltages of 80, 100, and 120 kVp were used in combination with five tube currents of 25, 50, 100, 200, and 400 mA; additionally, three slice thicknesses of 0.625, 1.25, and 2.5 mm and four reconstruction algorithms of adaptive statistical iterative reconstruction (ASIR-V) of 30, 60, and 90% were used. For each protocol, accuracy of the CAD system was evaluated for nine target GGNs of 6, 8, or 10 mm in size. The cut-off size was set to 5 mm to minimize false positives. Results Among the 180 combinations of tube voltage, tube current, slice thickness, and reconstruction algorithms, combination of 80 kVp, 200 mA, and 1.25-mm slice thickness with an ASIR-V of 90% had the best performance in the detection of GGNs with six true positives and no false positives. Other combinations had fewer than five true positives. In particular, any combinations with a 0.625-mm slice thickness had 0 true positive and at least one false positive result. Conclusion Low-voltage chest CT with a thin slice thickness and a high iterative reconstruction algorithm improve the detection rate of GGNs with a CAD system in a phantom model, and may have potential in lung cancer screening.
Cite
CITATION STYLE
Kang, S., Kim, T. H., Shin, J. M., Han, K., Kim, J. Y., Min, B., & Park, C. H. (2020). Optimization of a chest computed tomography protocol for detecting pure ground glass opacity nodules: A feasibility study with a computer-assisted detection system and a lung cancer screening phantom. PLoS ONE, 15(5). https://doi.org/10.1371/journal.pone.0232688
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.