Abstract
A radiation furnace that covers the temperature range from room temperature up to 1800 K has been designed and constructed for in situ synchrotron microtomography. The furnace operates under a vacuum or under any inert gas atmosphere. The two 1000 W halogen heating lamps are water- and air-cooled. The samples are located at the focus of these lamp reflectors on a rotary feedthrough that is connected to a driving rotation stage below the furnace. The X-ray beam penetrates the furnace through two X-ray-transparent vacuum-sealed windows. Further windows can be used for temperature control, sample changing and gas inflow and outflow. © 2009 International Union of Crystallography.
Author supplied keywords
Cite
CITATION STYLE
Grupp, R., Henkel, F., Nöthe, M., Banhart, J., Kieback, B., & Haibel, A. (2009). A 1800 K furnace designed for in situ synchrotron microtomography. Journal of Synchrotron Radiation, 16(4), 524–527. https://doi.org/10.1107/S0909049509020937