Diagnosis of Noise Inside Neonatal Incubators under Free-Field Conditions

5Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.

Abstract

The study aims to diagnose the sound pressure levels inside incubators in a controlled environment under free-field conditions. The tests were carried out in a semi-anechoic room under the standard UNE-EN ISO 3745:2012/A1:2018 in three different operating states: off, on, and on with a temperature alarm triggered. Sound pressure levels were analyzed in three different models of incubators, both inside and outside. The main noise indices analyzed were the corrected equivalent continuous level (LKeq) and the equivalent continuous level (Leq) in third-octave bands. The results obtained under normal operating conditions showed variations among the different incubators, with overall values between 48.8 and 56.3 dBA. The influence of the alarm considerably worsened these data. The values obtained showed that premature newborns are exposed to noise levels above international recommendations. All incubators tested showed the presence of tonal components, both outside and inside the incubator cabin, and, in some cases, low-frequency components, but no impulsivity components were observed in any case.

Cite

CITATION STYLE

APA

Fernández-Zacarías, F., Beira-Jiménez, J. L., Puyana-Romero, V., & Hernández-Molina, R. (2023). Diagnosis of Noise Inside Neonatal Incubators under Free-Field Conditions. Acoustics, 5(2), 354–366. https://doi.org/10.3390/acoustics5020021

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free