Abstract
Heat produced by a magnetic resonance (MR) imaging sequence in the vicinity of a conductive wire (pacemaker, electrodes, or catheter), is a subject of interest for the assessment of patient safety during imaging. For this purpose, the measurement of temperature rises during an MR imaging sequence using MR T1-based thermometry provides several advantages, mainly in its ability to retrieve in situ real-time thermal maps. Recent studies investigated the heat produced by an independent radiofrequency pulse, assessing MR imaging sequence heating using a specific MR thermometry sequence. This study focuses on self-heating for which the radiofrequency pulses used for measuring temperature create the heat. An experimental design was set up to evaluate T 1-based thermometry self-heating using a coupled/decoupled wire and to compare it with a reference temperature gathered by an optical fiber device. For the tested experimental set up, T1-based thermometry is in fairly good agreement with optical fiber reference temperature. © 2011 Wiley-Liss, Inc.
Author supplied keywords
Cite
CITATION STYLE
Detti, V., Grenier, D., Perrin, E., & Beuf, O. (2011). Assessment of radiofrequency self-heating around a metallic wire with MR T1-based thermometry. Magnetic Resonance in Medicine, 66(2), 448–455. https://doi.org/10.1002/mrm.22834