BACKGROUND The Pressure biofeedback unit (PBU) is an assessment tool used in clinical practice and research aimed to indirectly analyze the transversus abdominis (TrA) muscle activity. The concurrent validity of the PBU in a clinically relevant sample is still unclear. OBJECTIVE The purpose of this study was to evaluate the concurrent validity and diagnostic accuracy of the PBU in measuring TrA muscle activity in patients with chronic nonspecific low back pain. METHOD This study was performed using a validation, cross-sectional design. Fifty patients with chronic nonspecific low back pain were recruited for this study. To test the concurrent validity both PBU measures (index test) and superficial electromyographic measures (reference-standard test) were compared and collected by a physical therapist in a series of voluntary contraction maneuvers of TrA muscle. RESULTS Participants were on average 22 years old, weighed 63.7 kilos, 1.70 meters height and mean low back pain duration was 1.9 years. It was observed a weak and non-significant Phi coefficient (r=0.2, p<0.20). With regards to diagnostic accuracy tests, our results suggest a low sensitivity (60%) and specificity (60%) of the PBU. The positive predictive value was high (0.8) and negative predictive value was low (0.2). CONCLUSIONS Concurrent validity of the PBU in measuring TrA muscle activity in patients with chronic nonspecific low back pain is poor given the low correlation and diagnostic accuracy with superficial EMG.
CITATION STYLE
Lima, P. O. P., Oliveira, R. R., Moura Filho, A. G., Raposo, M. C. F., Costa, L. O. P., & Laurentino, G. E. C. (2012). Concurrent validity of the pressure biofeedback unit and surface electromyography in measuring transversus abdominis muscle activity in patients with chronic nonspecific low back pain. Brazilian Journal of Physical Therapy, 16(5), 389–395. https://doi.org/10.1590/s1413-35552012005000038
Mendeley helps you to discover research relevant for your work.