Comparison of Partial Digestion Procedures for Determination of Ca, Cr, Cu, Fe, K, Mg, Mn, Na, P, and Zn in Milk by Inductively Coupled Plasma-Optical Emission Spectrometry

31Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A fast procedure was developed for determination of Ca, Cr, Cu, Fe, K, Mg, Mn, Na, P, and Zn in milk samples. This procedure consisted of a partial digestion with hydrochloric acid on a hot plate. The results obtained were compared with 3 other digestion procedures (dry ashing and 2 microwave digestions). All the procedures showed similar precision levels, with coefficients of variation <10% for most analyzed elements. Accuracy was evaluated by using certified reference materials, and the values obtained were within the confidence intervals for these products. The results obtained were not considered statistically different. The partial digestion on a hot plate with HCI can be very practical for laboratories with relatively large numbers of sample analyses.

Cite

CITATION STYLE

APA

Kira, C. S., Maio, F. D., & Maihara, V. A. (2004). Comparison of Partial Digestion Procedures for Determination of Ca, Cr, Cu, Fe, K, Mg, Mn, Na, P, and Zn in Milk by Inductively Coupled Plasma-Optical Emission Spectrometry. Journal of AOAC International, 87(1), 151–156. https://doi.org/10.1093/jaoac/87.1.151

Register to see more suggestions