Simple Automated Verification of Field Size Indicator for Quality Assurance of Medical Linear Accelerator

  • Siti Hanan
  • Choirul Anam
  • Eko Hidayanto
N/ACitations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

Purpose: The purpose of this study is to automate the field size verification to facilitate mechanical check aspect medical linear accelerator (linac) quality assurance in a MATLAB-based algorithm on electronic portal imaging device (EPID) images. Methods: A total of 5 reference datasets (i.e. field sizes of 5 cm × 5 cm, 10 cm × 10 cm, 15 cm × 15 cm, 20 cm × 20 cm, and 25 cm × 25 cm) and 15 test datasets (i.e. reference field sizes plus 1 mm, 3 mm, and 5 mm increments) acquired by 6 MV Elekta Linac were used in this study. The proposed algorithm implemented a full automatic threshold with a value of 230 as a segmentation technique. The automated results were compared with manual results obtained using a ruler. Results: The automated results are comparable to manual results (i.e., the difference of both is within 2% or equal to 3 mm). The range of minimum to maximum difference between automated and manual was 0 - 3 mm and the maximum difference found in the 15.3 cm field size setting. Conclusions: We have successfully developed an automated procedure of field size verification and confirmed that the proposed algorithm provide a fast and accurate results.

Cite

CITATION STYLE

APA

Siti Hanan, Choirul Anam, & Eko Hidayanto. (2022). Simple Automated Verification of Field Size Indicator for Quality Assurance of Medical Linear Accelerator. International Journal of Scientific Research in Science and Technology, 55–60. https://doi.org/10.32628/ijsrst22922

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free