A differentially pumped harmonic filter on the Chemical Dynamics Beamline at the Advanced Light Source

101Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A differentially pumped rare gas cell has been developed to suppress undulator harmonics on the Chemical Dynamics Beamline at the Advanced Light Source. Greater than 104 suppression of the harmonics has been demonstrated with no measurable (<5%) attenuation of the fundamental. The overall design is presented, and vacuum and optical performance are reported. © 1995 American Institute of Physics.

Cite

CITATION STYLE

APA

Suits, A. G., Heimann, P., Yang, X., Evans, M., Hsu, C. W., Lu, K. T., … Kung, A. H. (1995). A differentially pumped harmonic filter on the Chemical Dynamics Beamline at the Advanced Light Source. Review of Scientific Instruments, 66(10), 4841–4844. https://doi.org/10.1063/1.1146161