Empirical two-point α-mixing model for calibrating the ECH 2O EC-5 soil moisture sensor in sands

105Citations
Citations of this article
122Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Recently improved ECH2O soil moisture sensors have received significant attention in many field and laboratory applications. Focusing on the EC-5 sensor, a simple and robust calibration method is proposed. The sensor-to-sensor variability in the readings (analog-to-digital converter (ADC) counts) among 30 EC-5 sensors was relatively small but not negligible. A large number of ADC counts were taken under various volumetric water contents (θ) using four test sands. The proposed two-point α-mixing model, as well as linear and quadratic models, was fitted to the ADC - θ data. Unlike for conventional TDR measurements, the effect of sensor characteristics is lumped into the empirical parameter α in the two-point a-mixing model. The value of α was fitted to be 2.5, yielding a nearly identical calibration curve to the quadratic model. Errors in θ associated with the sensor-to-sensor variability for the two-point α-mixing model were ±0.005 cm3 cm-3 for dry sand and ±0.028 cm3 cm-3 for saturated sand. In the validation experiments, the highest accuracy in water content estimation was achieved when sensor-specific ADCdry and ADCsat were used in the two-point α-mixing model. Assuming that α = 2.5 is valid for most mineral soils, the two-point α-mixing model only requires the measurement of two extreme ADC counts in dry and saturated soils. Sensor-specific ADC dry and ADCsat counts are readily measured in most cases. Therefore, the two-point α-mixing model (with α = 2.5) can be considered as a quick, easy, and robust method for calibrating the ECH 2O EC-5 sensor. Although further investigation is needed, the two-point α-mixing model may also be applied to calibrating other sensors.

Cite

CITATION STYLE

APA

Sakaki, T., Limsuwat, A., Smits, K. M., & Illangasekare, T. H. (2008). Empirical two-point α-mixing model for calibrating the ECH 2O EC-5 soil moisture sensor in sands. Water Resources Research, 44(4). https://doi.org/10.1029/2008WR006870

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free