Proton ion beam therapy demands for high resolution dosimetry due to the high dose gradients present in lateral confinement and final Bragg-peak. In polymer gels the reduction of the linear dose response in the area of the Bragg-peak is reported (Bragg-peak quenching), which is assumed to be mainly due to the high linear energy transfer (LET). We here investigate the impact of the spatial resolution in T2-mapping for accurate Magnetic Resonance Imaging (MRI)-based polymer gel dosimetry in the Bragg-peak for monoenergetic ion beams. We implemented MR-protocols for T2-mapping at microscopic resolution on a High-Field 7T human MR-scanner using an insert gradient system and sensitive rf-coils. The best results are obtained for an optimzed polymer gel based on THPC with an optimized MR-protocol for reduced measurement time and sufficient SNR at 0,547 mm pixel size. The dose in the fine Bragg-peak could be measured correctly for a monoenergetic proton beam as confirmed by Monte Carlo dose simulations. Such high spatial resolutions at minimum are necessary for an accurate measurement of the dose in the sharp Bragg-peak for monoenergetic ion beams. We demonstrate that at higher pixel size the dose levels may be underestimated due to spatial averaging in MRI-based polymer gel dosimetry.
CITATION STYLE
Berg, A., Wieland, M., Naumann, J., & Jaekel, O. (2013). High resolution dosimetry in monoenergetic proton beam therapy on a normoxic polymer gel: The importance of high spatial resolution for reduced Bragg-Peak-quenching. In Journal of Physics: Conference Series (Vol. 444). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/444/1/012054