Characterization of lipid-rich plaques using spectroscopic optical coherence tomography

  • Nam H
  • Song J
  • Jang S
  • et al.
18Citations
Citations of this article
38Readers
Mendeley users who have this article in their library.

Abstract

Intravascular optical coherence tomography (IV-OCT) is a high-resolution imaging method used to visualize the internal structures of walls of coronary arteries in vivo. However, accurate characterization of atherosclerotic plaques with gray-scale IV-OCT images is often limited by various intrinsic artifacts. In this study, we present an algorithm for characterizing lipid-rich plaques with a spectroscopic OCT technique based on a Gaussian center of mass (GCOM) metric. The GCOM metric, which reflects the absorbance properties of lipids, was validated using a lipid phantom. In addition, the proposed characterization method was successfully demonstrated in vivo using an atherosclerotic rabbit model and was found to have a sensitivity and specificity of 94.3% and 76.7% for lipid classification, respectively.

Cite

CITATION STYLE

APA

Nam, H. S., Song, J. W., Jang, S.-J., Lee, J. J., Oh, W.-Y., Kim, J. W., & Yoo, H. (2016). Characterization of lipid-rich plaques using spectroscopic optical coherence tomography. Journal of Biomedical Optics, 21(7), 075004. https://doi.org/10.1117/1.jbo.21.7.075004