Quantitative imaging characteristics of zirconium-89 on Gemini Time-Of-Flight PET/CT

1Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Purpose Interest in PET imaging using zirconium-89 (89 Zr) (t 1/2 =78.41 h)-labeled tracers for the tracking and quantification of monoclonal antibodies (mAbs) is growing, mainly because of its well-matched physical half-life with the biological half-life of intact mAbs. This study aims to evaluate the imaging characteristics of 89 Zr-PET in comparison with those obtained using fluorine-18 fluorodeoxyglucose (18 F-FDG) PET (gold standard tracer in PET imaging) using a Time-Of-Flight (TOF) PET/computed tomography (CT) scanner. Materials and methods The system's spatial resolution, sensitivity, scatter fraction (SF), image uniformity, and image quality were measured on a Gemini TOF PET/CT scanner according to the NEMA NU2-2001 protocols. The NEMA 2001 kit was used to carry out these measurements. Timing and energy resolutions were measured using 22 Na and 18 F-FDG point sources only. Results Spatial resolution in transverse and axial planes measured at 10 mm off access were 4.7 and 4.6 mm for 89 Zr and 18 F-FDG, respectively. At 100 mm, radial, tangential, and axial spatial resolution values were 5.2, 5.1, and 5.2 mm for 89 Zr and 5.1, 4.9, and 5.2 mm for 18 F-FDG, respectively. Sensitivity measured at the center of the field of view was 14.6 and 4.16 cps/kBq for 89 Zr and 18 F-FDG, respectively. SF was 32.6% for 89 Zr in comparison with 31.8% for 18 F-FDG. Image contrast for 89 Zr-PET images was 36.9 and 29.7% for 18 F-FDG for the smallest (10 mm)-sized sphere, and it was 70.6 and 72.8% for 89 Zr and 18 F-FDG, respectively, for the largest (37 mm)-sized sphere. Background variation was 10.3% for 89 Zr and 6.8% for 18 F-FDG for the smallest-sized sphere and 3.4 and 3.8% for 89 Zr and 18 F-FDG, respectively, for the largest-sized sphere. Conclusion In this study, we measured imaging characteristics of 89 Zr on a Gemini TOF PET/CT scanner. Our results show that 89 Zr has lower spatial resolution and noise-equivalent count rate with increased SF and background variation; however, it offered superior sensitivity and improved image contrast in comparison with 18 F-FDG. 89 Zr is an ideal radiotracer for immuno-PET imaging because of its physical half-life, which is well matched with mAbs, in addition to its affinity to be trapped inside the target cell after internalization of the mAbs.

Cite

CITATION STYLE

APA

Alanazi, S. F., Alzimami, K. S., Ghannam, M. M., Aljammaz, I. J., Alrumayan, F., & Sassi, S. A. (2016). Quantitative imaging characteristics of zirconium-89 on Gemini Time-Of-Flight PET/CT. Nuclear Medicine Communications, 37(12), 1238–1245. https://doi.org/10.1097/MNM.0000000000000602

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free