A simple, inexpensive, and clean methodology employing a combined spot test/diffuse reflectance spectroscopy procedure was developed and applied for the quality control analysis of lead in hair dyes, using a filter paper platform. The method is based on the complexation reaction of lead, present in a pH 2.8 buffer medium containing tartaric acid and sodium and potassium tartrate, with sodium rhodizonate (ROD) in an aqueous medium containing sodium dodecyl sulfate (SDS) surfactant, yielding a colored compound on the surface of a filter paper. The concentrations of ROD and SDS were optimized using experimental design. The linear range obtained for reflectometric measurements carried out at 545 nm was 4.70-140 mg L-1 of lead (R = 0.998), and the limits of detection and quantification were 1.12 and 3.75 mg L-1 , respectively. The proposed method was successfully applied to the analysis of lead in progressive hair dyes.

Cite

CITATION STYLE

APA

Luiz, V. H. M., Saraiva, L. M., Martins, Â. P., Pezza, L., & Pezza, H. R. (2015). Rapid determination of lead in progressive hair dye lotion by spot test/diffuse reflectance spectroscopy with a paper platform. Journal of the Brazilian Chemical Society, 26(10), 2137–2143. https://doi.org/10.5935/0103-5053.20150200