Damage to the cuff of EMG tube at endotracheal intubation by using a lightwand

6Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

Electromyogpraphic endotracheal tube (EMG tube) is a new device used to monitor recurrent laryngeal nerve integrity during thyroid surgery. The EMG tube has 2 pairs of electrodes on the surface of silicon-based tube reached to inner space of tube cuff. We experienced an unusual endotracheal tube-related problem from the distinct structural feature of the EMG tube. In this case, we intubated a patient who had difficult airway with the EMG tube using a lightwand. After successful endotracheal intubation, we could not expand the pilot balloon and ventilate the patient effectively. We removed the EMG tube and found that one of electrodes of the EMG tube is bended and made a right angle with the long axis of the tube, and perforated the tube cuff. So we report this case to make anesthesia providers aware that much more attention is needed to use EMG tube during endotracheal intubation. Copyright © 2010 Korean Society of Anesthesiologists.

Cite

CITATION STYLE

APA

Kim, H. S., Park, K. S., Kang, M. H., & Park, C. D. (2010). Damage to the cuff of EMG tube at endotracheal intubation by using a lightwand. Korean Journal of Anesthesiology, 59(SUPPL.). https://doi.org/10.4097/kjae.2010.59.S.S17