Abstract
OBJECTIVE - To characterize glucose levels during daily living using continuous glucose monitors (CGMs) in nondiabetic individuals. RESEARCH DESIGN AND METHODS - Seventy-four healthy children, adolescents, and adults aged 9-65 years with normal glucose tolerance used a blinded CGM device for 3 to 7 days. RESULTS - Sensor glucose concentrations were 71-120 mg/dl for 91% of the day. Sensor values were ≤60 or >140 mg/dl for only 0.2% and 0.4% of the day, respectively. Sensor glucose concentrations were slightly higher in children than adults (P = 0.009) and were slightly lower during the night than day (95 vs. 99 mg/dl, P < 0.001). CONCLUSIONS - Glucose values ≤60 and >140 mg/dl, measured with CGM, are uncommon in healthy, nondiabetic individuals. CGM may be useful to evaluate glucose tolerance in nondiabetic individuals over time. Furthermore, these data provide a basis for comparison for studies that use CGM to assess glucose control in subjects with diabetes. © 2010 by the American Diabetes Association.
Cite
CITATION STYLE
Fox, L. A., Beck, R. W., Xing, D., Chase, H. P., Gilliam, L. K., Hirsch, I., … Wilson, D. M. (2010). Variation of interstitial glucose measurements assessed by continuous glucose monitors in healthy, nondiabetic individuals. Diabetes Care, 33(6), 1297–1299. https://doi.org/10.2337/dc09-1971