Screen Printed Thick Film Reference Electrodes for Electrochemical Sensing

46Citations
Citations of this article
60Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

This paper presents printed thick film AgAgClKCl reference electrodes (RE) for electrochemical sensors. The screen-printed REs with 10-\mu \text{m} thick glass-KCl salt matrix layer exhibit a stable potential of 5 mV. Cyclic voltammetric analysis shows that the anodic and cathodic peak current of the RE increases with the scan rate in the range of 25-150 mVs-1. The analytical performance of the REs shows a stable open circuit potential for the NaCl concentrations in the range of 30-100 mM. Testing the presented REs for electrochemical pH sensor application (with RuO2-based sensitive electrode) the sensitivity of 55 mV/pH was noted in the pH range of 4.5-9. Evaluating the effect of temperature on the performance of REs, a potential variation of-3.8 mV/ °C was observed. Finally, a LabVIEW interface was developed to store, analyze, and calculate the sensitivity of the sensor under different temperature conditions. The LabVIEW interface can also be used to calculate the pH-value/temperature of unknown solutions under known temperature/pH conditions.

Cite

CITATION STYLE

APA

Manjakkal, L., Vilouras, A., & Dahiya, R. (2018). Screen Printed Thick Film Reference Electrodes for Electrochemical Sensing. IEEE Sensors Journal, 18(19), 7779–7785. https://doi.org/10.1109/JSEN.2018.2840349