Patient-centric microsampling for abrocitinib pharmacokinetics: multiple-analytes assay bridging using Tasso device

9Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Aim: The feasibility of using Tasso devices (Tasso-SST® and Tasso+) collecting capillary blood samples for measuring abrocitinib and its metabolites were evaluated, and assay concordance established between capillary and venous blood samplings. Methods: Capillary serum and venous plasma concentrations were measured using their respective qualified and validated assays. Concentration and exposure comparisons were conducted for abrocitinib and its metabolites (M1, M2 and M4) to establish assay concordance. Results: The correlation coefficient between capillary serum and venous plasma concentrations were >0.98 for all four analytes from three separate assays, and PK parameters (AUClast and Cmax) were compared and met bioequivalence criteria. Conclusion: These results demonstrate the feasibility of patient-centric microsampling device, such as Tasso, in future abrocitinib pediatric study.

Cite

CITATION STYLE

APA

Tripathy, S., Wan, K. X., Shahin, M. H., Winton, J. A., Malhotra, B. K., & Kavetska, O. (2024). Patient-centric microsampling for abrocitinib pharmacokinetics: multiple-analytes assay bridging using Tasso device. Bioanalysis. Taylor and Francis Ltd. https://doi.org/10.1080/17576180.2024.2388939