Comparison of GeneXpert MTB/RIF assay and LED-FM microscopy for the diagnosis of extra pulmonary tuberculosis in Khyber Pakhtunkhwa, Pakistan

22Citations
Citations of this article
73Readers
Mendeley users who have this article in their library.
Get full text

Abstract

GeneXpert is one of the recent technological instruments used to diagnose tuberculosis in a short span of time. In this study, the performance of GeneXpert MTB/RIF assay for the diagnosis of extrapulmonary tuberculosis (EPTB) was compared with light-emitting diode Fluorescent Microscopy (LED-FM) in Khyber Pakhtunkhwa, Pakistan. A total of 737 EPTB samples were collected from tuberculosis (TB) suspected patients. Out of these samples, male to female ratio was 53% (n = 390) to 47% (n = 347) respectively. The sensitivity and specificity was 73% and 100% for GeneXpert, while 40% and 100% for LED-FM microscopy. This shows that the sensitivity of GeneXpert is 40–50%, higher than LED-FM microscopy. GeneXpert also detected low number of bacilli as compared to LED-FM microscopy.

Cite

CITATION STYLE

APA

Khan, A. S., Ali, S., Khan, M. T., Ahmed, S., Khattak, Y., Abduljabbar, … Sajjad, W. (2018). Comparison of GeneXpert MTB/RIF assay and LED-FM microscopy for the diagnosis of extra pulmonary tuberculosis in Khyber Pakhtunkhwa, Pakistan. Brazilian Journal of Microbiology, 49(4), 909–913. https://doi.org/10.1016/j.bjm.2018.02.011