Abstract
We investigated the deterioration in plasticity of the multi-use Portex Tracheal Guide ('bougie') with repeated sterilisation. Six bougies were prepared by washing them between 0 and 100 times (in accordance with the manufacturer's guidelines). Two tests were employed: a bench test in which rapid serial photographs were taken of the bougies uncoiling from a preformed curve; and a manikin-based test in which the six bougies were used in a simulated difficult airway. The bench test demonstrated a progressive deterioration in plasticity with repeated washing. However, the manikin-based test showed no significant difference between bougies in the incidence of oesophageal placement (p = 0.74). Time to placement differed significantly only between the two most-washed bougies but was broadly similar. We suggest therefore that the manufacturer's limit of five washings may be unnecessarily cautious. You can respond to this article at © 2011 The Authors. Anaesthesia © 2011 The Association of Anaesthetists of Great Britain and Ireland.
Cite
CITATION STYLE
Dawes, T. J. W., & Ford, P. N. R. (2011). The effect of sterilisation on the plasticity of multi-use Eschmann gum elastic bougies: A bench and manikin study. Anaesthesia, 66(12), 1134–1139. https://doi.org/10.1111/j.1365-2044.2011.06897.x