Abstract
Background: MR elastography (MRE) is an emerging technique for tumor diagnosis. MRE actuation devices require precise mechanical design and radiofrequency engineering to achieve the required mechanical vibration performance and MR compatibility. Method: A method of designing a general-purpose, compact and inexpensive MRE actuator is presented. It comprises piezoelectric bimorphs arranged in a resonant structure designed to operate at its resonant frequency for maximum vibration amplitude. An analytical model was established to understand the device vibration characteristics. Results: The model-predicted performance was validated in experiments, showing its accuracy in predicting the actuator resonant frequency with an error < 4%. The device MRI compatibility was shown to cause minimal interference to a 1.5 tesla MRI scanner, with maximum signal-to-noise ratio reduction of 7.8% and generated artefact of 7.9 mm in MR images. Conclusions: A piezoelectric MRE actuator is proposed, and its implementation, vibration issues and future work are discussed. © 2011 John Wiley & Sons, Ltd.
Author supplied keywords
Cite
CITATION STYLE
Tse, Z. T. H., Chan, Y. J., Janssen, H., Hamed, A., Young, I., & Lamperth, M. (2011). Piezoelectric actuator design for MR elastography: Implementation and vibration issues. International Journal of Medical Robotics and Computer Assisted Surgery, 7(3), 353–360. https://doi.org/10.1002/rcs.405