Abstract
Objectives. We have developed a pulsed xenon ultraviolet light-based real-time air disinfection system with rapid and effective disinfection by using high-intensity pulse germicidal UV. Disinfection of the ambulance's environment is critical in the prevention of infectious cross contamination. Methods. In this study, a pulsed xenon ultraviolet light-based air disinfection system was established for real-time air disinfection in ambulances. In this system, a pulsed xenon ultraviolet (PX-UV) was used to generate broad-spectrum (200-320 nm), high-intensity ultraviolet light to deactivate and kill bacteria and viruses. The results showed that the use of PX-UV could be effective in reducing E. coli, Staphylococcus albus, and environmental pathogens level in ambulances (≥90% reduction in 30 mins). Results. This device was relatively simple and easy to use and does not leave chemical residues or risk exposing patients and workers to toxic chemicals. Conclusions. This appears to be a practical alternative technology to achieve automated air disinfection in ambulances.
Cite
CITATION STYLE
Song, L., Li, W., He, J., Li, L., Li, T., Gu, D., & Tang, H. (2020). Development of a Pulsed Xenon Ultraviolet Disinfection Device for Real-Time Air Disinfection in Ambulances. Journal of Healthcare Engineering, 2020. https://doi.org/10.1155/2020/6053065