A comparison of two calibration methods for ultrasonic hydrophones

13Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Although miniature ultrasonic hydrophones are frequently used to measure the acoustic pressure distributions from diagnostic ultrasound sources, relatively little attention has been devoted to the methods for absolute calibration of these hydrophones. In this study a polyvinylidene (PVDF) hydrophone was used to compare two calibration methods currently in use. One is based on a reciprocity technique and the second involves the planar scanning of a source transducer having a known radiated ultrasonic power. The reciprocity method revealed that the hydrophone response did not vary by more than ± 1.6dB from -262.8dB re 1V/μPa over the frequency range of 1-10 MHz. For the planar scanning technique seven ultrasound sources between 1-10 MHz were used, and all calibration points were within ±0.5 dB of the corresponding points found by the method of reciprocity. © 1982.

Cite

CITATION STYLE

APA

Gloersen, W. B., Harris, G. R., Stewart, H. F., & Lewin, P. A. (1982). A comparison of two calibration methods for ultrasonic hydrophones. Ultrasound in Medicine and Biology, 8(5), 545–548. https://doi.org/10.1016/0301-5629(82)90086-2