Abstract
The relative yield of 3γ to 2γ annihilation was measured for a positron emitter, 22Na, with the new scintillator detectors lanthanum chloride (LaCl3:Ce) and lanthanum bromide (LaBr3:Ce), which had been characterised for comparison with high-purity germanium (HPGe) and sodium iodide (NaI(Tl)) detectors. The information obtained from the ortho-positronium 3γ decay in positron emission tomography (PET) can be a measure of the oxygen content in biological tissues by determination of this relative yield. However, it requires high resolution spectroscopy and detection efficiency. Characterisation of the new generation of scintillator detectors determines whether they could replace conventional scintillators and semiconductors. A series of experiments was carried out with different samples in order to study the effect of orthopositronium formation. The peak-to-peak and the peakto- valley methods were compared in the measurement of the relative yield of 3γ to 2γ annihilation. © Akadémiai Kiadó, Budapest, Hungary 2011.
Author supplied keywords
Cite
CITATION STYLE
Alkhorayef, M., Alzimami, K., Alfuraih, A., Alnafea, M., & Spyrou, N. M. (2012). Measurement of three gamma annihilation by lanthanum-based crystals compared with NaI(Tl) and HPGe. Journal of Radioanalytical and Nuclear Chemistry, 291(2), 493–496. https://doi.org/10.1007/s10967-011-1316-3