Variability of Retinal Thickness Measurements in Tilted or Stretched Optical Coherence Tomography Images

  • Uji A
  • Abdelfattah N
  • Boyer D
  • et al.
N/ACitations
Citations of this article
18Readers
Mendeley users who have this article in their library.

Abstract

PURPOSE: To investigate the level of inaccuracy of retinal thickness measurements in tilted and axially stretched optical coherence tomography (OCT) images. METHODS: A consecutive series of 50 eyes of 50 patients with age-related macular degeneration were included in this study, and Cirrus HD-OCT images through the foveal center were used for the analysis. The foveal thickness was measured in three ways: (1) parallel to the orientation of the A-scan (Tx), (2) perpendicular to the retinal pigment epithelium (RPE) surface in the instrument-displayed aspect ratio image (Ty), and (3) thickness measured perpendicular to the RPE surface in a native aspect ratio image (Tz). Mathematical modeling was performed to estimate the measurement error. RESULTS: The measurement error was larger in tilted images with a greater angle of tilt. In the simulation, with axial stretching by a factor of 2, Ty/Tz ratio was > 1.05 at a tilt angle between 13 degrees to 18 degrees and 72 degrees to 77 degrees , > 1.10 at a tilt angle between 19 degrees to 31 degrees and 59 degrees to 71 degrees , and > 1.20 at an angle ranging from 32 degrees to 58 degrees . Of note with even more axial stretching, the Ty/Tz ratio is even larger. Tx/Tz ratio was smaller than the Ty/Tz ratio at angles ranging from 0 degrees to 54 degrees . The actual patient data showed good agreement with the simulation. The Ty/Tz ratio was greater than 1.05 (5% error) at angles ranging from 13 degrees to 18 degrees and 72 degrees to 77 degrees , greater than 1.10 (10% error) angles ranging from 19 degrees to 31 degrees and 59 degrees to 71 degrees , and greater than 1.20 (20% error) angles ranging from 32 degrees to 58 degrees in the images axially stretched by a factor of 2 (b/a = 2), which is typical of most OCT instrument displays. CONCLUSIONS: Retinal thickness measurements obtained perpendicular to the RPE surface were overestimated when using tilted and axially stretched OCT images. TRANSLATIONAL RELEVANCE: If accurate measurements are to be obtained, images with a native aspect ratio similar to microscopy must be used.

Cite

CITATION STYLE

APA

Uji, A., Abdelfattah, N. S., Boyer, D. S., Balasubramanian, S., Lei, J., & Sadda, S. R. (2017). Variability of Retinal Thickness Measurements in Tilted or Stretched Optical Coherence Tomography Images. Translational Vision Science & Technology, 6(2), 1. https://doi.org/10.1167/tvst.6.2.1

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free