Development and characterisation of a high-sensitivity X-ray CT polymer gel dosimeter

1Citations
Citations of this article
2Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Polymer gel dosimeters have shown potential for clinical 3D dosimetry; however, their use has been limited due to low sensitivity and reliance on scarcely available magnetic resonance imaging. This study aimed to optimise a PASSAG (Poly AMPS Sodium Salt And Gelatin) polymer gel dosimeter for X-ray computed tomography, to enhance its clinical feasibility. The total monomer concentration was increased to improve sensitivity, and different cosolvents were tested to enhance the limited solubility of N, N’-methylenebisacrylamide, the crosslinker. n-propanol was identified as the optimal cosolvent, allowing for an 18.3% monomer concentration, 30% crosslinker to total comonomer mass gel, at a 3:7 cosolvent-to-water ratio. The optimised formulation, PASSAG-N (PASSAG– n-propanol), consisted of 54.4% w/w deionised water, 23.3% n-propanol, 12.8% 2-acrylamido-2-methylpropane sulfonic acid sodium salt, 5.5% N, N’-methylenebisacrylamide, 4.0% gelatin, and 0.089% (4.65mM) tetrakis (hydroxymethyl)phosphonium chloride. The dosimeter was irradiated within a standard timeframe to assess its sensitivity, and theoretical calculations confirmed its equivalence to water, soft tissue, brain, and muscle. Compared to a cosolvent-free formulation, PASSAG-N exhibited a 250% increase in Hounsfield unit (HU) change, demonstrating enhanced sensitivity. The optimised gel showed a linear response over a 1–12 Gy dose range, with an average sensitivity of 1.072 ± 0.041 HU Gy⁻¹ and a dose resolution ≤ 0.31 Gy, making it a promising alternative for clinical X-ray computed tomography-based dosimetry. This study highlights the potential of PASSAG-N as a highly sensitive and potentially practical polymer gel dosimeter for clinical applications.

Cite

CITATION STYLE

APA

Lumley, J. L., Rowshanfarzad, P., Ibrahim, M., Djukelic, M., & Henry, D. J. (2025). Development and characterisation of a high-sensitivity X-ray CT polymer gel dosimeter. Physical and Engineering Sciences in Medicine, 48(3), 1311–1323. https://doi.org/10.1007/s13246-025-01586-2

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free