COMP Report: CPQR technical quality control guidelines for CyberKnife® Technology

18Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. This particular TQC contains detailed performance objectives and safety criteria for CyberKnife® Technology. The quality control recommendations in this document are based upon previously published guidelines and the collective experience of all Canadian sites using this technology. This TQC guideline has been field tested at the newest Canadian CyberKnife installation site and includes recommendations for quality control of the Iris™ and InCise™ MLC collimation systems.

Cite

CITATION STYLE

APA

Vandervoort, E., Patrocinio, H., Chow, T., Soisson, E., & Nadeau, D. B. (2018). COMP Report: CPQR technical quality control guidelines for CyberKnife® Technology. Journal of Applied Clinical Medical Physics, 19(2), 29–34. https://doi.org/10.1002/acm2.12263