Abstract
In this paper we report on the development and validation of two different methods for posaconazole quantification from serum samples, HPLC/UV and bioassay. Both methods have been validated according to international guidelines and were also applied to the analysis of 61 trough serum samples from treated patients. A good correlation between both methods was observed. The HPLC method, more laborious and expensive, was demonstrated to be more accurate, precise and faster (analytical range 0.125-16μg/mL, accuracy between -2.48 and 3.70% and precision between 2.77 and 5.93%, with an analytical run time of 11min), making it a valuable tool for reference laboratories that centralize high numbers of samples. The microbiological method, however, is simple and offers sufficient precision and accuracy (analytical range 0.125-16μg/mL, accuracy between -8.10 and 3.77% and a precision between 4.52 and 10.07%), to be used to monitor posaconazole. It may be a valid alternative to chromatographic methods in clinical laboratories without specialized facilities. © 2011 The Authors. Clinical Microbiology and Infection © 2011 European Society of Clinical Microbiology and Infectious Diseases.
Author supplied keywords
Cite
CITATION STYLE
Cendejas-Bueno, E., Forastiero, A., Rodriguez-Tudela, J. L., Cuenca-Estrella, M., & Gomez-Lopez, A. (2012). HPLC/UV or bioassay: Two valid methods for posaconazole quantification in human serum samples. Clinical Microbiology and Infection, 18(12), 1229–1235. https://doi.org/10.1111/j.1469-0691.2011.03732.x