Abstract
Purpose: To improve vessel visibility in time-of-flight MR angiography (TOF-MRA) by careful consideration of coil choice, coil position, and frequency offset and profile of the nonspatlally selective chemical shift selective (CHESS) pre-saturation pulse. Materials and Methods: The effects of both the CHESS and the excitation radiofrequency (RF) pulses on flow signal and signals from stationary substances were evaluated by changing the spatial area where RF pulses were applied to upstream flow in a flow phantom and in human subjects. The difference between the eight-channel phased-array receive-only coil and the transmit-receive coil was evaluated. Results: The CHESS pulse suppresses the flow signal over a wider frequency range than the signals from stationary substances, especially when using the body coil for transmission. Even without presaturation pulse, the excitation pulse slightly suppressed the flow signal. Adjusting the position of the transmit-receive coil relative to the head improved these TOF-MRA images. The results were better than those obtained with the eight-channel coil. Conclusion: The excitation and the nonspatlally selective CHESS pulses degraded the flow signal. Our results suggest that reduced spatial extent of RF pulse application to upstream flow can improve image quality of TOF-MRA. This result can be implemented on conventional scanners. © 2008 Wiley-Liss, Inc.
Author supplied keywords
Cite
CITATION STYLE
Shonai, T., Carpenter, J. S., Lemieux, S. K., Harada, K., Omori, K., Kaneko, N., & Fukushima, T. (2008). Improvement of vessel visibility in time-of-flight MR angiography of the brain. Journal of Magnetic Resonance Imaging, 27(6), 1362–1370. https://doi.org/10.1002/jmri.21371
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.