Utility of matrix-assisted laser desorption ionization-time of flight mass spectrometry following introduction for routine laboratory bacterial identification

168Citations
Citations of this article
105Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) was evaluated prospectively in a diagnostic laboratory. Nine hundred twenty-seven organisms were tested in triplicate; 2,351/2,781 (85%) species and 2,681/2,781 (96%) genus identifications were correct. Known issues such as the misidentification of alpha-hemolytic streptococci as Streptococcus pneumoniae were easily corrected. Identifications cost AUD$0.45 per isolate and were available in minutes. MALDI-TOF MS is rapid, accurate, and inexpensive. Copyright © 2011, American Society for Microbiology. All Rights Reserved.

Cite

CITATION STYLE

APA

Neville, S. A., LeCordier, A., Ziochos, H., Chater, M. J., Gosbell, I. B., Maley, M. W., & Van Hal, S. J. (2011). Utility of matrix-assisted laser desorption ionization-time of flight mass spectrometry following introduction for routine laboratory bacterial identification. Journal of Clinical Microbiology, 49(8), 2980–2984. https://doi.org/10.1128/JCM.00431-11