Abstract
Aim: To design an accommodating intraocular lens with extended accommodative range that can be adapted to current standard phacoemulsification and endocapsular implantation technique. Method: Ray tracing analysis and lens design; cadaver eye implantation. Results: Ray tracing analysis indicated that axial movement of an exaggerated converging anterior optic linked by spring loaded haptics to a compensatory static diverging posterior optic produced greater change in conjugation power of the eye compared to axial movement of a single optic lens. A dual optic one piece foldable silicone lens was constructed and implanted via a 4 mm corneal incision into the capsular bag of two cadaver eyes. Conclusion: A dual optic intraocular lens design can increase the optical effect of a given displacement and suggests improvements for accommodating intraocular lenses.
Cite
CITATION STYLE
McLeod, S. D., Portney, V., & Ting, A. (2003). A dual optic accommodating foldable intraocular lens. British Journal of Ophthalmology, 87(9), 1083–1085. https://doi.org/10.1136/bjo.87.9.1083